Overview
This device is a multifunctional electrical measurement instrument calibration system designed in accordance with the requirements of JJG124-2005 “Verification Regulations for Ammeters, Voltmeters, Power Meters, and Ohmmeters,” JJG 596-1999, and JJG 307-2006. The core technology of the device's meter section employs a high-precision power frequency AC signal acquisition system composed of a digital signal processor (DSP) and a 16-bit high-speed analog-to-digital converter. The signal source section utilizes a controllable sine wave and distorted waveform signal generator built with a DSP and a 16-bit high-speed digital-to-analog converter.
The device features high accuracy, stable and reliable operation, and convenient and flexible handling.
2.1 Capable of semi-automatic or manual verification of fundamental errors in various power frequency meters within electrical systems (voltmeters, ammeters, wattmeters, frequency meters, power factor meters, phase meters), as well as alterations caused by influencing factors such as voltage, current, waveform, and power factor.
2.2 The power supply section can generate distorted waves containing harmonics from the 2nd to the 31st order. The number of harmonics, their order, amplitude, and phase relative to the fundamental wave can all be controlled via programming.
2.3 The power amplifier operates within a bandwidth of 40Hz to 1kHz with excellent linearity. The current amplifier functions as a constant current source, while the voltage amplifier serves as a constant voltage source. Due to its lightweight design, this device is particularly suitable for field calibration applications.
2.4 Equipped with a USB interface. Through host computer software (optional), the device can be controlled by a computer to perform automatic or manual testing, with results processed and managed.
2.5 Equipped with a large-capacity non-volatile memory, it can store raw test data from 300 calibrated meters for review and upload.
2.6 Capable of calibrating electricity meters at load points or user-selected points.
3.1 AC Voltage Range: 50V, 100V, 200V, 400V, 600V; Maximum Output Capacity: 20VA;
3.2 AC Current Range: 0.5A, 1A, 2.5A, 5A, 10A, 20A; Maximum Output Capacity: 20VA;
3.3 AC Voltage and Current Adjustment Range: 0–130% FS; Adjustment Resolution: 5×10⁻⁵;
3.4 Power Frequency AC Voltage and Current Accuracy: 0.05% FS;
3.5 Active Power (50V–400V) Accuracy: 0.05% FS;
3.6 Reactive Power (50V–400V) Accuracy 0.1% FS;
3.7 Current Phase Accuracy Relative to Same-Name Phase Voltage 0.05°;
3.8 Frequency Adjustment Range 45–65Hz, Adjustment Resolution 0.001Hz, Setpoint Accuracy 0.01Hz;
3.9 Phase Adjustment range 0 to 359.99°, Adjustment resolution 0.01°;
Power factor Adjustment range 0 to ±1, Adjustment resolution 0.0001; Setpoint accuracy 0.0005;
3.10 AC Voltage/Current Output Waveform Distortion ≤0.3%;
3.11 AC Voltage/Current and Power Output Stability ≤0.01% FS /60s;
3.12 Harmonics 2nd to 31st order, amplitude 0–40%, phase adjustment per harmonic 0.01°·N (N is harmonic order);
3.13 DC Voltage Ranges: 75mV, 75V, 150V, 300V, 500V, 1000V; Maximum Output Capacity: 20W;
3.14 DC Current Ranges: 0.5A, 1A, 2.5A, 5A, 10A, 20A; Maximum Output Capacity: 20W;
3.15 DC Voltage and Current Adjustment Range: 0–120% FS (0–105% FS for 1000V range), Adjustment Resolution: 5×10⁻⁵;
3.16 DC Voltage and Current Output Stability: ≤0.01% FS /60s (peak-to-peak);
3.17 DC voltage setpoint accuracy 0.05% FS;
3.18 DC current setpoint accuracy 0.1% FS;
3.19 DC 75mV output (load ≥5 Ω) accuracy 0.1% FS;
3.20 Energy Meter Measurement (50V–400V) Total Error 0.05% (Pulse Output: Three-Phase Rated Value = 6Hz);
3.21 Operating Power Supply Single-Phase 220V±10%, 50Hz±5%;
3.22 Operating Environment: Temperature 20°C ±10°C, Relative Humidity ≤85%RH;
3.23 Dimensions & Weight: Approx. 460×460×175mm³, 20kg (Deluxe chassis: 24kg).
We provide you with free consultation services
Engineers provide guidance for installation and commissioning
Technicians can provide on-site maintenance services and receive regular training
Perfect technical consulting service system